THURLBY THANDAR INSTRUMENTS TSX series High Current Laboratory Power Supplies More Info> METER SPECIFICATIONS Meter types: Separate4 digit meters for voltage and current with 12.5mm (0.5 ) LED displays. Meter resolutions: 10mV, 10mA. Meter accuracies: Voltage (0.2%+1 digit) Current (0.5%+1 digit). MECHANICAL & ENVIRONMENTAL Electrical safety: Complies with EN61010-1. EMC: Complies with EN50081-1 and EN50082-1. o o o Temperature: +5C to +40C operating, 20% to 80% RH, -40C to o +70C storage. Size: 210x 130x 350mm (WxHxD) (half rack widthx 3U height), optional rack mounting kit available. High power levels in a compact & lightweight casing Weight: 5.0kg (TSX versions). 35V-10A & 18V-20A models with more to come 5.5kg (TSX-P versions). Bench or rack mounting, front & rear terminals FRONT PANEL CONTROLS (standard versions) Voltage setting: Via single rotary controls for coarse and fine control. Very low noise, excellent transient reponse Current setting: Via single turn semi-logarithmic rotary control. CV & CI operation with automatic crossover Overvoltage setting: Via screwdriver adjustable preset potentiometer. Comprehensive protection including variable OVP trip Output On/Off: Via large paddle lever switch. High setting resolution, remote sense terminals FRONT PANEL CONTROLS (P versions) Voltage setting: Direct keyboard entry or quasi-analogue High accuracy digital meters, current meter damping rotary control. Keyboard/Rotary setting of all parameters (TSX-P) Current setting: Direct keyboard entry or quasi-analogue rotary control. Watts display, non-volatile storage of 25 settings (TSX-P) Overvoltage setting: Direct keyboard entry. Fully programmable with bus readback of V and I (TSX-P) Output On/Off: Push button with dual indicator lamps. GPIB (.2) & RS232 interfaces (TSX-P) Note: all voltage and current levels set via the keyboard are displayed on a separate 0.3 4 digit display. This entry preview system ensures that Model Range: the user can observe the value entered before it is effected thus avoiding TSX1820- Single output,0 to 18V at0 to 20A. possible error. The display is also used for setting additional functions TSX3510- Single output,0 to 35V at0 to 10A. and for displaying watts. TSX1820P-0 to 18V at0 to 20A, GPIB& RS-232 interfaces. Additional keyboard Increase or decrease voltage or current in TSX3510P-0 to 35V at0 to 10A, GPIB& RS-232 interfaces. functions: user-selectable steps (delta mode). Store/recall voltage, current& OVP levels from non-volatile memory (25 memories). OUTPUT SPECIFICATIONS Set digital interface type (RS232 or GPIB), set baud Operating modes: Constant voltage or constant current with automatic rate, set address. crossover. Voltage range: 0V to 35V (TSX3510/TSX3510P). DIGITAL INTERFACES (P versions) 0V to 18V (TSX1820/TSX1820P). RS232: Variable baud rate, 9600 baud maximum,9 pin D Current range: 0A to 10A (TSX3510/TSX3510P). connector (male). Fully compatible with standard 0A to 20A (TSX1820/TSX1820P). RS232 or TTi addressable RS232 system (ARC). Overvoltage protection: 10% to 110% of max. output voltage. IEEE-488 (GPIB): Conforming with IEEE488.1& IEEE488.2. Setting resolution: 10mV, 10mA. Operational functions: Set voltage set current set OVP set output On/Off Load regulation: <0.01% of max. O/P for 90% change. read output voltage/current. Line regulation: <0.01% of max. O/P for 10% change. Setting resolution: Voltage- 10mV Current- 10mA. Output impedance: <1m in constant voltage mode. Setting accuracy: Voltage- (0.1%+ 10mV) Current- (0.2%+ 20mA). >5k in constant current mode. Response times: Interface- <15ms (single command) Ripple& noise: <1mV RMS typical in constant voltage. PSU- Depends on Load conditions, typically 150ms <3mA RMS typical in constant current. to within 0.1% of final value (except for voltage re- HF common mode noise: Typically <3mV RMS, <10mV pk. duction with low load current which will be longer). Transient load response:<20us to within 50mV of set level for 90% load Readback resolution: Voltage- 10mV Current- 10mA. change. o Readback accuracy: Voltage- (0.1%+1 digit) Temperature coefficient: typically <100ppm/ C. Current- (0.5% +1 digit). Overvoltage protection Operating software: Software for operating the PSUs under GPIB or delay: <200us. RS232 control is available includinga Labwindows* Protection functions: Overvoltage trip, Regulator overtemperature driver and ARC-TALK software fora PC. Sense miswiring. Status indication: Output on/off lamp, Constant voltage mode lamp Thurlby Thandar Instruments Ltd. operates a policy of continuous development Constant current mode lamp, Trip message. and reserves the right to alter specifications without prior notice. Output switch: Electronic. Output terminals: 4mm output terminals at front, screw terminals for Designed and built in the EEC by: output and sense at rear. Output protection: Full forward and reverse protection via OVP and di- ode clamp. INPUT SPECIFICATIONS Input voltage range: 180V to 270V RMS, 90V to 135V RMS, 47 to 63Hz. Thurlby Thandar Instruments Ltd. Power requirement: 750VA max. Glebe Road, Huntingdon. Cambs. PE18 7DX England Voltage range selection: Rear panel slide switch. Tel: +44 (0)1480 412451 Fax: +44 (0)1480 450409 e-mail: sales ttinst.co.uk Web: X-ON Electronics Largest Supplier of Electrical and Electronic Components Click to view similar products for Bench Top Power Supplies category: Click to view products by TTI Aim manufacturer: Other Similar products are found below : NL200 PR20 ZUPNC403 Z60-7-L-U ZUPNC402 GDM-8342 GPIB PSW 250-4.5 CPX200DP AX-8450A TPM-3003 HMP2020 HMP2030 HMP4040 1320 1350 UT804 1405 1410 1513 1514 1513 1550 1651A 1665 1666 1667 SDP-2405-000G 1673 1693 1694 MX100TP 1737 1739 1762 1788 TPM-3005 1900 1902B 9174B GDM-8245 GDM-8255A GDM-8341 PSW 160-7.2 PSW 30-36 PSW 80-13.5 Z10-20-LAN-U PSW 30-72 QL355P SII Z36-6-LAN-U HCS-3400-USB