The MBC60 Series of open-frame medical power supplies, with its wide universal 90-264 VAC input range, is available at 60 W of output power and a variety of single and multiple output voltages. th The MBC series was designed to 4 edition medical approvals and provides 2x MOPP (Means of Patient Protection) isolation for Class I and Class II installations. These medical power supplies are ideal for, monitoring, home health equipment as well as surgical devices. 50 - 65 W Single to Triple Outputs Ultra High Efficiency > 85% Form Factor 4 x 2 x 1.2 inches (101.6 x 50.8 x 30.48 mm) Low leakage current < 250 A No Load Power < 0.3 W Approved to EN/IEC 60601-1 2x MOPP IEC Protection Class Options: o Class I: Earth pin J4 (no suffix) o Class II: No Earth pin (-2 suffix) Conducted EMI EN 55022-B, FCC Part 15 Level B Medical Safety Agency Approvals RoHS Compliant Cover Kit Accessory Available Dialysis Surgical Devices Monitoring Home Health Pumps Ultrasound 2 MBC60 Series 1 2 3 4 MODEL VOLTAGE MAX LOAD MIN LOAD RIPPLE POWER MBC60-1005G 5.0 V 10.0 A 0.0 A 1.5% 50 W MBC60-1012G 12 V 5.42 A 0.0 A 1% 65 W MBC60-1015G 15 V 4.33 A 0.0 A 1% 65 W MBC60-1024G 24 V 2.71 A 0.0 A 1% 65 W MBC60-1048G 48 V 1.35 A 0.0 A 1% 65 W V = 5.2 V 8.0 A 0.5 A 1.5% 1 MBC60-3000G V = 12.5 V 3.0 A 0.1 A 1% 60 W 2 V = -12.8 V 0.5 A 0.0 A 1% 3 V = 5.2 V 8.0 A 0.5 A 1.5% 1 MBC60-3001G V = 24 V 1.5 A 0.1 A 1% 60 W 2 V = -12.8 V 0.5 A 0.0 A 1% 3 V = 5.2 V 8.0 A 0.5 A 1.5% 1 MBC60-3002G V = 15 V 2.5 A 0.1 A 1% 60 W 2 V = -15 V 0.5 A 0.0 A 1% 3 V = 3.3 V 6.0 A 1.0 A 1.5% 1 MBC60-3003G V = 5 V 3.0 A 0.1 A 1% 45 W 2 V = -12.8 V 0.5 A 0.0 A 1% 3 5 COVER-60-XBC Metal cover kit accessory 1 Class II version available. Add suffix -2 at the end of the Model Number 2 Maximum current per output channel. Do not exceed total output power rating. 3 Min Load specified to meet cross regulation. 4 Ripple is peak to peak with 20 MHz bandwidth and 10 F (Electrolytic capacitor) in parallel with a 0.1 F capacitor at rated line voltage and load ranges. 5 When used in Cover Kit, de-rate output power to 70 % under all operating conditions. Specifications are for nominal input voltage, 25C unless otherwise stated. PARAMETER DESCRIPTION / CONDITION SPECIFICATION Input Voltage Universal 90 - 264 VAC 6 Input Frequency 47 to 400 Hz 120 VAC: 1.5 A max. Input Current 230 VAC: 0.75 A max. Single output models < 0.3 W No Load Power Multi output models < 0.5 W 120 VAC: 30 A max. Inrush Current 230 VAC: 60 A max. 120 VAC: < 140 A Leakage Current 230 VAC: <250 A Switching Frequency Typical 67 67 kHz 6 Safety approved: 47 to 63 Hz tech.support psbel.com