DC/DC Medical Converter THM 3 Series, 3 Watt Compact DIP-24 plastic case I/O isolation 5000 VAC rated for 250 VAC working voltage IEC 60601-1 certification for 2 x MOPP Risk management process according to ES 60601-1 IEC 60601-1 ISO 14971 incl. risk management file UL 62368-1 IEC 62368-1 Acceptance criteria for electronic The THM 3 series is a range of medical 3 Watt DC/DC converters in DIP-24 assemblies acc. to IPC-A-610 Level 3 plastic package with wide 2:1 input voltage range. They provide a reinforced Low leakage current <2 A isolation system for 5000 VAC and a very low leakage current of less than 2 A. The units are approved to IEC/EN/ES 60601-1 3rd ed. for 2 x MOPP and Operating temperature 40C to 90C come along with an ISO 14971 risk management file. Design and production conform to the quality management system ISO 13485. With a high efficiency EMC compliance to IEC 60601-1-2 4th of up to 88% and highest grade components the converters can reliably oper- edition and EN55032 class A ate in an ambient temperature range of 40C up to +90C. They constitute a 5-year product warranty reliable solution not only for medical equipment but also for demanding ranges of application such as transportation, control & measurement or IGBT drivers. Models Order Code Input Voltage Output 1 Output 2 Efficiency Range Vnom Imax Vnom Imax typ. THM 3-0510 3.3 VDC 1 000 mA 81 % THM 3-0511 5 VDC 600 mA 85 % THM 3-0512 12 VDC 250 mA 86 % 4.5 - 9 VDC THM 3-0513 15 VDC 200 mA 88 % (5 VDC nom.) THM 3-0515 24 VDC 125 mA 86 % THM 3-0521 +5 VDC 300 mA 5 VDC 300 mA 83 % THM 3-0522 +12 VDC 125 mA 12 VDC 125 mA 86 % THM 3-0523 +15 VDC 100 mA 15 VDC 100 mA 86 % THM 3-1210 3.3 VDC 1 000 mA 82 % THM 3-1211 5 VDC 600 mA 85 % THM 3-1212 12 VDC 250 mA 87 % 9 - 18 VDC THM 3-1213 15 VDC 200 mA 87 % (12 VDC nom.) THM 3-1215 24 VDC 125 mA 87 % THM 3-1221 +5 VDC 300 mA 5 VDC 300 mA 84 % +12 VDC 125 mA 12 VDC 125 mA 88 % THM 3-1222 THM 3-1223 +15 VDC 100 mA 15 VDC 100 mA 87 % THM 3-2410 3.3 VDC 1 000 mA 82 % THM 3-2411 5 VDC 600 mA 85 % THM 3-2412 12 VDC 250 mA 87 % 18 - 36 VDC THM 3-2413 15 VDC 200 mA 87 % (24 VDC nom.) THM 3-2415 24 VDC 125 mA 87 % THM 3-2421 +5 VDC 300 mA 5 VDC 300 mA 83 % THM 3-2422 +12 VDC 125 mA 12 VDC 125 mA 87 % THM 3-2423 +15 VDC 100 mA 15 VDC 100 mA 86 % THM 3-4810 3.3 VDC 1 000 mA 81 % THM 3-4811 5 VDC 600 mA 84 % THM 3-4812 12 VDC 250 mA 87 % 36 - 75 VDC THM 3-4813 15 VDC 200 mA 87 % (48 VDC nom.) THM 3-4815 24 VDC 125 mA 87 % THM 3-4821 +5 VDC 300 mA 5 VDC 300 mA 83 % THM 3-4822 +12 VDC 125 mA 12 VDC 125 mA 86 % THM 3-4823 +15 VDC 100 mA 15 VDC 100 mA 86 % Page 1 / 6 www.tracopower.com January 6, 2021THM 3 Series, 3 Watt Options - Optional models with alternative pinning on demand - Optional models with adjustable output (backorder with MOQ - Optional models with remote-control function non stocking item) - Optional models with adjustable output and remote-control function Input Specifications Input Current - At no load 5 Vin models: 20 mA typ. 12 Vin models: 10 mA typ. 24 Vin models: 6 mA typ. 48 Vin models: 4 mA typ. Surge Voltage 5 Vin models: 16 VDC max. (3 s max.) 12 Vin models: 25 VDC max. (3 s max.) 24 Vin models: 50 VDC max. (3 s max.) 48 Vin models: 100 VDC max. (3 s max.) Under Voltage Lockout 5 Vin models: 3 VDC min. / 4 VDC typ. / 4.4 VDC max. 12 Vin models: 7 VDC min. / 8 VDC typ. / 8.8 VDC max. 24 Vin models: 15 VDC min. / 16 VDC typ. / 17.5 VDC max. 48 Vin models: 31.5 VDC min. / 33 VDC typ. / 34.5 VDC max. Recommended Input Fuse 5 Vin models: 1 600 mA (slow blow) 12 Vin models: 800 mA (slow blow) 24 Vin models: 500 mA (slow blow) 48 Vin models: 315 mA (slow blow) (The need of an external fuse has to be assessed in the final application.) Input Filter Internal Pi-Type Output Specifications Output Voltage Adjustment 10% to +20% (15 & 24 Vout single models) 10% (other models) (Only for optional models with adjustable output) (By external trim resistor) See application note: www.tracopower.com/overview/thm3 Output power must not exceed rated power Voltage Set Accuracy 1% max. Regulation 0.2% max. - Input Variation (Vmin - Vmax) single output models: dual output models: 0.5% max. - Load Variation (0 - 100%) single output models: 0.2% max. dual output models: 1% max. (Output 1) 1% max. (Output 2) 5% max. - Cross Regulation dual output models: (25% / 100% asym. load) Ripple and Noise - single output 3.3 Vout models: 30 mVp-p typ. (w/ 10 F X7R) (20 MHz Bandwidth) 5 Vout models: 30 mVp-p typ. (w/ 10 F X7R) 12 Vout models: 40 mVp-p typ. (w/ 10 F X7R) 15 Vout models: 40 mVp-p typ. (w/ 10 F X7R) 24 Vout models: 50 mVp-p typ. (w/ 4.7 F X7R) - dual output 5 / -5 Vout models: 30 / 30 mVp-p typ. (w/ 10 F X7R) 12 / -12 Vout models: 40 / 40 mVp-p typ. (w/ 10 F X7R) 15 / -15 Vout models: 40 / 40 mVp-p typ. (w/ 10 F X7R) All specifications valid at nominal voltage, full load and +25C after warm-up time unless otherwise stated. Page 2 / 6 www.tracopower.com January 6, 2021