TS318-11C55 Thermopile Sensor SPECIFICATIONS Thermopile IR-Sensor For Contactless Temperature Measurement Single Element Small Package for Ear Thermometer High Signal Flat Filter Accurate Reference Sensor Thermopiles are mainly used for contactless temperature measurement in many applications. Their function is to transfer the heat radiation emitted from the objects into a voltage output. SENSOR SOLUTIONS ///TS318-11C55 Page 1 09/2015 TS318-11C55 Thermopile Sensor FEATURES High Signal NTC Reference Sensor Small TO-18 Package 5.5m Long Wave Pass Filter APPLICATIONS Medical Ear Thermometer ABSOLUTE MAXIMUM RATINGS Parameter Symbol Min Typical Max Unit Description Storage Temperature TS -20 +20 +85 C permanent Storage Temperature TS -20 +20 +100 C non permanent PERFORMANCE SPECS Parameter Symbol Value Unit Condition Operating Ambient Temperature TAmb -20 to +85 C permanent Operating Ambient Temperature TAmb -20 to +100 C non permanent Package TO-18 2 Absorber Area A 0.8 0.8 mm Thermopile Resistance R 70 30 T = +25C TP k Amb Temperature Coefficient TCRTP -0.06 0.04 %/K TAmb = +25C to +75C of Thermopile Resistance TAmb = +25C, TObj = +100C, Voltage Response VTP 8.4 2.1 mV DC, totally filled field of view Temperature Coefficient TCVTP -0.45 0.08 %/K TAmb = +25C to +75C of Voltage Response Noise Equivalent Voltage NEV 34 nV/Hz TAmb = +25C Rise Time 12 5 ms 63 Ambient Temperature Sensor NTC Ambient Temperature Sensor R 100 5 T = +25C NTC k Amb Resistance Beta Value of NTC -Value 3955 0.3% K TAmb = 0C to +50C 10/2015 SENSOR SOLUTIONS /// TS318-11C55 Page 2